Specified low-power radio equipment (Item 8 of Article 2 Paragraph 1 of Certification Ordinance)
for Medical telemeter Type D and F


Medical telemeter Type A, B, C


Test items Technical standards
Medical telemeter
Type D
Medical telemeter
Type E
Class of emissions F7D, F8D, F9D, G7D

FM-FM, GMSK
F7D, F8D, F9D, G7D

PWM-FM, GMSK
Assigned frequency or designated frequency A part of 420 - 450MHz band A part of 420 - 450MHz band
Number of channels and channel separation 54
100kHz
12
Tolerance of frequency
(×10-6)
20 20
Tolerance of occupied bandwidth 64kHz 320kHz
Tolerance of spurious and unwanted emission intensity 2.5µW 2.5µW
Antenna power Designated value 1mW or less 10mW or less
Tolerance +20%, -50% +20%, -50%
Tolerance of adjacent channel leakage power -40dB
(100 kHz separation, 64 kHz bandwidth)
-40dB
(500 kHz separation, 320 kHz bandwidth)
Limit of secondary radiated emissions 4nW 4nW
Other equipment or function
-------
-------
Note
-------
-------

Notice: Regulations in the Ordinance Regulating Radio Equipment shall have priority over the descriptions in the sheet when any difference is found between them. Further above descriptions do not cover all of the regulations.

Copyright 2006 TELECOM ENGINEERING CENTER.