Specified low-power radio equipment (Item 8 of Article 2 Paragraph 1 of Certification Ordinance)
for Medical telemeter Type A,B and C


Medical telemeter Type D, F


Test items

Technical standards

Medical telemeter
Type A
Medical telemeter
Type B
Medical telemeter
Type C
Class of emissions F1D, F2D, F3D, F7D, F8D, F9D

FM - FM, Binary FSK
F7D, F8D, F9D


FM - FM, Binary FSK
F7D, F8D, F9D, G7D


FM - FM, GMSK
Assigned frequency or designated frequency A part of 420 - 450MHz band A part of 420 - 450MHz band A part of 420 - 450MHz band
Number of channels and channel separation 480
12.5kHz
234
25kHz
114
50kHz
Tolerance of frequency
(×10-6)
4 (OBW : 4 - 8.5kHz)
10 (OBW : 4 kHz or less)
4 (OBW : 12 - 16kHz)
10 (OBW : 12kHz or less)
10
Tolerance of occupied bandwidth 8.5kHz 16kHz 32kHz
Tolerance of spurious and unwanted emission intensity 2.5µW 2.5µW 2.5µW
Antenna power Designated value 1mW or less 1mW or less 1mW or less
Tolerance +20%, -50% +20%, -50% +20%, -50%
Tolerance of adjacent channel leakage power -40dB
(12.5kHz separation, 8.5kHz bandwidth)
-40dB
(25kHz separation, 16kHz bandwidth)
-40dB
(50kHz separation, 32kHz bandwidth)
Limit of secondary radiated emissions 4nW 4nW 4nW
Other equipment or function
-------
-------
-------
Note OBW : Occupied bandwidth
-------

Notice: Regulations in the Ordinance Regulating Radio Equipment shall have priority over the descriptions in the sheet when any difference is found between them. Further above descriptions do not cover all of the regulations.